Configuration Reference

All defaults live in flimkit/configs.py and can be overridden via CLI args or the GUI.

Fitting Defaults

Parameter Default Description
Tau_min 0.145 ns Lower lifetime bound
Tau_max 45.0 ns Upper lifetime bound
n_exp 3 Number of exponential components
D_mode 'both' Fitting mode: 'summed', 'perPixel', or 'both'
binning_factor 1 Spatial binning for per-pixel fitting
Optimizer 'de' 'de' (Differential Evolution) or 'lm_multistart'
MIN_PHOTONS_PERPIX 10 Minimum photons for per-pixel fitting
OUT_NAME 'flim_out' Default output prefix

Phasor Filtering Defaults

Parameter Default Description
PHASOR_FILTER None Filter method: 'gaussian', 'median', 'wavelet', or None
PHASOR_FILTER_SIGMA 1.0 Gaussian σ in pixels
PHASOR_FILTER_SIZE 3 Median filter kernel size (pixels)
PHASOR_FILTER_WAVELET 'db4' Wavelet family for wavelet denoising
PHASOR_FILTER_LEVEL 1 Wavelet decomposition level

Optimiser Settings

Parameter Default Description
lm_restarts 8 Levenberg-Marquardt multi-start restarts
de_population 30 DE population size
de_maxiter 5000 DE maximum iterations
n_workers -1 (source) / 1 (compiled) CPU cores for DE; capped at 1 in the compiled app to avoid multiprocessing issues
de_n_starts None Independent DE runs per summed fit. None means 1
de_seed 42 Seed of the first DE start. Start k uses de_seed + k

Tri-exponential accuracy and DE starts

Up to 0.14 a tri-exponential summed fit landed somewhere different for each differential-evolution seed. On 10 synthetic tri-exponential decays at 10^6 photons, fitted 128 times each with only the seed changed, single runs were off by anything from 0.02 % to 83 %. The error of a run is the median over its three lifetimes of |τ_fit / τ_true - 1|.

Lifetime error of each of 128 DE runs on each of 10 tri-exponential decays, with the median per decay

Keeping the lowest polished cost of several seeds helped, and the gain flattened out after about 12. The table gives the median over the 10 decays from 3000 random draws of N seeds.

Seeds kept by lowest cost (N) Median error 10th-90th percentile
1 6.4 % 3.6-10.2 %
12 2.6 % 1.8-3.8 %
64 2.4 % 1.9-2.8 %

Median lifetime error against the number of seeds kept by lowest cost, with the 10th to 90th percentile band

Each run's polished cost above the best run for the same decay against its lifetime error

The scatter turned out to come from the Levenberg-Marquardt polish after DE, not from several separate minima. It worked on the raw parameters, lifetimes of about 10^-9 s beside amplitudes of about 10^5, and stopped before it reached the bottom. Starting the same polish from the true lifetimes reached a lower deviance than DE on all 10 decays. From 0.15.0 the polish works on log10 of the lifetimes with ten times the evaluation budget, and DE then reaches the same deviance as the start from the truth on all 10. Every seed now lands on the same fit, so one start is the default again.

3-exp, 10^6 photons, paired IRF Long τ Mid τ Short τ Time per fit
FLIMKit 0.15.0, 1 start 0.4 % 2.1 % 3.3 % 5 s
FLIMKit 0.15.0, 12 starts 0.4 % 2.1 % 3.3 % 41 s
FLIMKit 0.14.1 0.3 % 6.7 % 7.4 %
LAS X 0.4 % 2.8 % 3.3 %
FLIMJ (MLE) 0.7 % 5.1 % 7.3 %
SymPhoTime 64 0.6 % 7.8 % 8.6 %
SPCImage 14.9 % 127 % 55 %

The values are medians over the same 10 decays, all fitted with the same measured IRF. Times are for one core. For one- and two-exponential decays 0.14.1 was already as close to the truth as LAS X or closer, for example 0.7 % against 1.4 % on the short lifetime of a two-exponential decay. The remaining error at 10^6 photons is two to five times the Cramér-Rao bound for these decays (0.1, 0.7 and 1.4 % for the long, mid and short lifetimes at 1σ) and doesn't come from the optimiser, because the start from the truth lands in the same place. I have not found where it comes from.

More starts can still be asked for, with --n-starts on the command line, DE starts in Expert Fit Settings, or n_starts= on fit_summed and fit_summed_tail. Start k uses seed --seed + k, and the run with the lowest polished Poisson cost is kept. Bootstrap refits always use one start. The summary records n_starts, start_costs (the polished cost of each start, None where the polish failed), n_near_best (starts within 10^-3 relative of the best cost) and the winning seed. If a second start ever reaches a lower cost than the first, that is worth reporting.

DE updates its population the same way whatever the number of workers, so a fit with the same seed gives the same result on one core or many. Before 0.15.0 a single-core fit used a different update rule, so single-core results moved slightly.

Low photon counts

Below about 10^5 photons a three-exponential decay doesn't hold enough information to pin down three lifetimes, whatever the fitting tool. The Cramér-Rao bound gives the smallest scatter any unbiased fit can reach. For the same 10 decays, at 1σ:

Photons Long τ Mid τ Short τ
10^6 0.1 % 0.7 % 1.4 %
10^4 1.1 % 6.5 % 13.9 %
10^3 3.7 % 21 % 44 %

Holding every lifetime of every one of these decays within 10 % takes about 45,000 photons. Low counts are still FLIMKit's weak spot for three exponentials, as they are for every tool: 0.15.0 changes what is reported, not what the photons can resolve. Below that the fit surface is a flat valley. A fit lands near wherever it started, so a fit that starts near the expected lifetimes, for example from values typed in by hand, can look better than DE without fitting the data any better.

FLIMKit reports what the data can support. When any lifetime of a two- or three-exponential fit is uncertain by more than 10 % (from the bootstrap if it ran, otherwise from the curvature), the summary sets components_resolved to false, and the terminal, the Fit Summary and the ROI result lead with the mean lifetimes and mark the components as not resolved. The mean lifetimes are well determined even when the components are not. On the same 10 decays:

What is reported 10^4 photons 10^3 photons
Three free lifetimes, short τ 33 % off 76 % off
Free fit, intensity-weighted mean lifetime 0.5 % off 1.9 % off
Free fit, amplitude-weighted mean lifetime 3.0 % off 5.8 % off
Lifetimes fixed from a fit with more photons, worst fraction 4.2 points off 4.1 points off

The not-resolved check flagged all 10 decays at both counts. To get the fractions of a small region, fix the lifetimes from a fit with more photons, usually the whole image: tick Fix lifetimes from the image fit in the ROI and phasor-cursor fit options, or pass fixed_taus_ns= to fit_summed. Only the amplitudes, the background and the IRF shift are then fitted, and the summary sets fixed_taus. Per-pixel fits already work this way.

Display Range Settings

Pixel values outside the range are clamped to the boundary, not zeroed.

Parameter Default Description
TAU_DISPLAY_MIN None Min lifetime (ns) for tau images
TAU_DISPLAY_MAX None Max lifetime (ns) for tau images
INTENSITY_DISPLAY_MIN None Min photon count for intensity images
INTENSITY_DISPLAY_MAX None Max photon count for intensity images

Machine IRF Settings

Parameter Default Description
MACHINE_IRF_DIR flimkit/machine_irf (source) / ~/.flimkit/machine_irf (compiled) Storage directory
MACHINE_IRF_DEFAULT_PATH The Preferences choice if set, else the user copy if present, else the bundled default Resolved at startup, and again when Preferences are saved
MACHINE_IRF_ALIGN_ANCHOR 'peak' Alignment landmark during IRF construction
MACHINE_IRF_REDUCER 'median' Aggregation method across paired IRFs
MACHINE_IRF_FIT_STRATEGY 'fixed' Runtime fitting strategy
MACHINE_IRF_FIT_BG True Fit background offset
MACHINE_IRF_FIT_SIGMA False Fit Gaussian broadening
MACHINE_IRF_FIT_TAIL False Fit exponential tail

Cost Functions

Function Description
poisson Poisson deviance (C-statistic). Recommended
chi2 Neyman chi-squared (legacy)

Fit Diagnostics

The cost function above selects the optimizer objective. The fields below are post-fit diagnostics and do not change how the model is fitted.

The existing reduced_chi2_pearson, reduced_chi2_tail_pearson, and per-pixel chi2_r fields are retained for compatibility with historical Leica LAS X comparisons. Their one-count model floor means they are not generally expected to equal one for sparse decays.

calibrated_chi2_pearson, calibrated_chi2_tail_pearson, and the per-pixel calibrated_chi2_r map divide the same residual sum by its expected contribution under a fixed Poisson model:

$$Q_{\mathrm{cal}} = \frac{\sum_i (Y_i-m_i)^2/\max(m_i,1)}{\sum_i \min(m_i,1)}.$$

The fixed-model expectation is one. Parameter fitting and data-selected windows can introduce a smaller additional shift, so this diagnostic is not a classical chi-square p-value.

rise_chi2_per_bin_pearson is the Pearson chi-squared per bin over the fitted bins before the tail window, which covers the rising edge and the peak. The terminal and the Fit Summary show it next to the tail value. When the IRF is the wrong width or in the wrong place, the misfit sits on the rising edge, so this number climbs well above the tail value while the tail stays close to one. On a synthetic decay fitted with an IRF 2.7 times too wide, it was more than five times the tail value.

Chi-squared and photon count

A reduced chi-squared that grows with the photon count doesn't on its own mean the data are faulty. A small error in the IRF or the model leaves the same fractional misfit in every bin. The misfit grows in proportion to the counts, and the Poisson noise only grows with their square root, so the misfit shows at high counts and hides at low ones. On a Leica STELLARIS rhodamine 6G measurement, pulse-picked to 19.5 MHz so that no decay is left over from the previous pulse, the reduced deviance was 2.1, 5.5 and 19.7 at 0.4, 1.7 and 6.6 million photons.

Two things help find where the misfit comes from. If rise_chi2_per_bin_pearson is much larger than the tail value, look at the IRF first. If the residuals before the rising edge follow the end of the decay, look at incomplete decay.

Parameter uncertainties

Every summed fit (reconvolution, tail and distribution) reports a 1σ standard error for each lifetime, amplitude and amplitude fraction, both mean lifetimes, the IRF shift and the fitted background, in fields ending _err. They come from the curvature of the fit at its solution. The model's derivatives over the fit window, weighted by the Poisson variance of each bin, give the Fisher information, and its inverse is the covariance. A parameter sitting on one of its bounds is treated as fixed and has no error. tau_corr holds the correlation between the lifetimes.

On ATTO488 (2.46 million photons) the tail fit gives 4.187 ± 0.0034 ns, where SymPhoTime 64's bootstrap gives 4.189 ± 0.0029 ns. Over 20 simulated 1-exponential decays (τ = 2.5 ns, 200,000 photons) the predicted error was 0.0061 ns and the spread of the fitted values was 0.0057 ns. For 2-exponential decays (3.0 and 0.8 ns) the prediction was about 1.4 times the spread, so it errs large there.

These are local estimates. They can't see a different solution elsewhere that fits just as well, so for a 3-exponential fit they are a lower bound on the real uncertainty.

The bootstrap doesn't have that problem. It draws a new Poisson count for every bin of the measured decay, refits with the same settings and fit window, and repeats (25 times by default, seeded so the same data gives the same errors). The spread of the refitted lifetimes is the error, and the 16th and 84th percentiles give a range that doesn't have to be symmetric. The summary holds them as taus_ns_boot_err, taus_ns_boot_lo and taus_ns_boot_hi (tau_centers_ns_... for a distribution fit). The GUI, fit_cli.py and the guided terminal UI run it by default for single FOV, stitched and ROI fits. From Python it is off unless you pass bootstrap=25 to fit_summed, fit_summed_tail or fit_summed_dist. Batch, timelapse and z-stack fits don't run it.

Each bootstrap refit runs DE once, even when the fit itself used several starts (see Tri-exponential accuracy and DE starts).

For a synthetic 3-exponential decay (4.06, 1.36 and 0.71 ns, one million photons) fitted 16 times with fresh noise, τ2 scattered by 51% and τ3 by 24%. On one of those decays the curvature errors were 5.8% and 26%, and the bootstrap gave 24% and 33%, with both true values inside its 16th to 84th percentile range. On a 1-exponential decay the two agree (both 0.1%), and on ATTO488 the bootstrap gives ±0.0038 ns against SymPhoTime 64's ±0.0029 ns.

A lifetime with more than 10% relative error, or two lifetimes correlated beyond |r| = 0.95, adds a ⚠ row to the Fit Summary and a warning to the terminal output. When the bootstrap has run, these use its errors and correlations.

A fit with two or more lifetimes also warns when one of them is shorter than three times the IRF FWHM and carries less than 5% of the photons (its amplitude times its lifetime, as a share of the total). An IRF that doesn't match the data often produces a component like this, for example about 1% of the photons at about 200 ps beside a single-exponential dye. The test uses the photon share because a fast component can hold a large share of the amplitude and still very few photons: 16% of the amplitude at 0.25 ns beside 3.85 ns is 1.2% of the photons. The warning sits in fit_warnings as well as uncertainty_warnings, and is kept when the bootstrap replaces the others. It doesn't mean the component is false, only that the IRF should be checked before it is read as part of the sample.

Incomplete decay

At a high repetition rate a long lifetime hasn't finished decaying when the next pulse arrives. The leftover shows as a raised baseline before the rising edge that follows the end of the previous decay. A lifetime of 4 ns at 80 MHz (12.5 ns between pulses) leaves about 4% of its amplitude behind.

Pass the pulse period in ns with --rep-period-ns (12.5 for 80 MHz), or auto to read it from the file, set Incomplete decay, pulse period (ns) in Expert Fit Settings, or pass rep_period_ns= to fit_summed, fit_summed_tail, fit_summed_dist, fit_per_pixel or fit_per_pixel_dist. The fit then adds the decay from every earlier pulse, and the summary records rep_period_ns. In the CLI, the GUI and the batch fits, the per-pixel fit uses the same period as the summed fit. auto reads the period PTU and B&H files record; a file that doesn't record one is refused with a message. The period must be at least as long as the histogram, give or take one bin, because Leica histograms are rounded to whole bins (133 bins of 97 ps is 12.90 ns, for a 12.82 ns period). A period shorter than that is refused, because more than one pulse would land in the window.

What it changes depends on the histogram. The model convolves over the histogram with an FFT, which wraps the end of the window round to the start. When the histogram covers exactly one pulse period, as it usually does on a Leica FALCON, the wrap already carries the leftover from earlier pulses, so the lifetimes come out the same with or without the term. What changes is the amplitudes. With the term on they're per pulse, where without it each one is inflated by 1 / (1 - exp(-T/τ)), so the long components are overweighted. On a synthetic 4 and 0.8 ns decay at 50/50, folded from a train of pulses 12.5 ns apart, the fractions were 0.507 and 0.493 without the term and 0.496 and 0.504 with it. When the histogram is shorter than the period, the wrap is wrong and the term matters for the lifetimes too: with a 10 ns histogram of the same train and a 4 ns lifetime, the fit gave 3.90 ns at a reduced chi-squared of 10.4 without the term, and 3.96 ns at 0.76 with it.

Each kind of fit handles it its own way:

The earlier projection for fixed tau (--estimator nnls) subtracts a background taken from the bins before the peak. With incomplete decay those bins hold the end of the previous decay, not background, so that estimator stays biased with the term on. The default Poisson fit, which fits the offset, doesn't have this problem.

Real data agree. Three Leica STELLARIS rhodamine 6G measurements at 19.5 MHz (0.4, 1.7 and 6.6 million photons) were folded into a 12.8 ns period to emulate 78 MHz. Unfolded, they fit at 3.92 ns. Folded, the one-exponential fit gave 3.78 to 3.81 ns with the term off and moved by no more than 0.013 ns with it on, so the remaining 3% doesn't come from the leftover decay. On those fits rise_chi2_per_bin_pearson was about eight times the full-window value, which points at the IRF.

Folding the same files into 1/2, 1/3, 1/4 and 1/5 of the 51.3 ns period (39 to 97 MHz) shortens the lifetime steadily, to 3.71 ns at 97 MHz on all three photon levels, with or without the term, which moves it by 0.013 ns at most. Cutting the fit window of the unfolded data to the same lengths shortens it by only 1%, to 3.88 ns at 10 ns. I cannot account for the rest. A mismatched IRF combining with the fold is the likely cause, but that has not been tested.

A fit whose model cuts the decay off at the end of the window instead of wrapping it gets about 3.33 ns on the same data. That 15% shortfall comes from the model, not the data.

Checking an instrument calibration

A single-component standard with a known lifetime shows most calibration problems. Use a photon rate low enough to rule out pile-up, and acquire it on every detector you use, one after another.

A short standard tests the IRF more directly than a long one. With a lifetime of 1 ns or less, almost nothing is left from the previous pulse even at 80 MHz, so a high chi-squared or a fast component can't come from incomplete decay and points at the IRF or the timing calibration. Erythrosin B in water, at about 90 ps, is one such standard.